Relevance of Monitoring Atrial Fibrillation in Clinical Practice

Register or Login to View PDF Permissions
Permissions× For commercial reprint enquiries please contact Springer Healthcare: ReprintsWarehouse@springernature.com.

For permissions and non-commercial reprint enquiries, please visit Copyright.com to start a request.

For author reprints, please email rob.barclay@radcliffe-group.com.
Average (ratings)
No ratings
Your rating

Abstract

The monitoring of atrial fibrillation (AF) is performed using a variety of tools, ranging from the conventional Holter electrocardiogram to modern implantable loop recording with remote data exchange. The main clinical areas in AF where monitoring is crucial for decision-making are catheter and surgical ablation, as well as anticoagulation to prevent strokes. Identifying the patient cohort at risk – e.g., those with subclinical silent AF – is a challenge. In addition, the interaction of AF with implanted devices – e.g. AF-triggered inadequate shock therapy – should be the object of continuous monitoring. The prevention of inadequate shock delivery in particular is of major clinical importance.

Disclosure:The authors have no conflicts of interest to declare.

Received:

Accepted:

Support:The publication of this article was funded by Medtronic. The views and opinions expressed are those of the authors and not necessarily those of Medtronic.

Acknowledgements:The authors would like to thank Mirko de Melis from the Medtronic Bakken Research Centre and Alphons Vincent from Medtronic Europe for their support.

Correspondence Details:Thorsten Lewalter, Head, Department of Cardiology and Intensive Care, Isar Heart Centre Munich, Sonnenstr. 24–26, 80331 Munich, Germany. E: thorsten.lewalter@isarkliniken.de

Copyright Statement:

The copyright in this work belongs to Radcliffe Medical Media. Only articles clearly marked with the CC BY-NC logo are published with the Creative Commons by Attribution Licence. The CC BY-NC option was not available for Radcliffe journals before 1 January 2019. Articles marked ‘Open Access’ but not marked ‘CC BY-NC’ are made freely accessible at the time of publication but are subject to standard copyright law regarding reproduction and distribution. Permission is required for reuse of this content.

The monitoring of atrial fibrillation (AF) can be performed using a great variety of strategies and tools. Strategies range from monitoring only symptomatic AF (e.g., post-catheter ablation with or without surface electrocardiogram [ECG] documentation) to continuously monitoring heart rhythm using implantable pacemakers, implantable cardioverter defibrillators (ICDs) or subcutaneous implantable cardiac monitors (ICMs).1,2 With implanted devices, AF monitoring can be performed by interrogation during clinic visits as well as remotely.

Recent studies using different monitoring tools, ranging from transtelephonic monitoring to pacemakers, have indicated that the correlation between AF episodes and patient symptoms is very poor: many AF episodes are asymptomatic, whereas many AF-like symptoms are not related to AF episodes.3–13

The relevance of AF monitoring in clinical practice includes evaluation of the efficacy of a rhythm control strategy such as AF ablation and identification of the need for oral anticoagulation.14,15 In ICD patients, AF occurrence may increase the risk of inappropriate shocks and should therefore prompt careful evaluation of concomitant drug therapy and device programming. In cardiac resynchronisation therapy (CRT) patients, AF with fast ventricular response may limit the efficacy of cardiac resynchronisation and AF ablation or atrioventricular (AV) node ablation may be indicated.16

Current Status of Monitoring Tools

There are a variety of monitoring tools available to detect and record occurrences of AF. In clinical practice, because of the rather poor sensitivity of single standard ECG for paroxysmal AF, 24-hour ECG recording (Holter) is often used, allowing the detection of previously unrecognised AF.17–19

The extension of the ECG recording time from 24 to 72 hours increases the prevalence of paroxysmal AF after stroke from 1.2 to 6.1 %, suggesting that the probability of detecting AF increases with the duration of heart rhythm monitoring.20–22 External loop recorders (ELRs) are used to monitor the heart rhythm for one week or longer in an ambulatory patient. ELRs perform an event-triggered (e.g., onset of AF) or patient-triggered (e.g., onset of palpitations) sampling of the ECG during the period of use, with a limited storage capacity – for example 20 minutes.22–24 Other implantable devices capable of intracardiac atrial electrogram recording, such as dual-chamber pacemakers and defibrillators, can detect AF more appropriately due to their continuous monitoring capability.25 It was demonstrated that, by minimising the risk of artefact detection caused by myo-potentials or other sources of electrical interference, the detection of atrial tachyarrhythmia (AT) episodes lasting at least five minutes is well correlated with a proven diagnosis of AF.26 Several clinical studies have been successfully conducted using the diagnostic features of implanted dual chamber pacemakers or ICDs to evaluate AF burden and assess different therapeutic strategies.27–30

Another option for AF monitoring consists of subcutaneous ICMs, which, unlike pacemakers and ICDs, cannot sense endocardial atrial activity, but rely on the analysis of consecutive RR intervals to diagnose AF. The irregularity of the RR interval is now a proven parameter for AF detection, based on mathematical tools such as the Lorenz plot. In a Lorenz plot, each RR interval is plotted against the previous value of the RR interval, and this can be displayed graphically and used to discriminate between AF and sinus rhythm. A validation study, the Reveal XT performance trial (XPECT), showed that an ICM equipped with an algorithm for AF detection can accurately measure AF burden (98.5 %) and is very sensitive (96.4 %) to identify patients with AF, independent of symptoms. The device automatically stores a two-minute ECG strip for visual inspection and confirmation by the physician.31,32

Intermittent Versus Continuous Atrial Fibrillation Monitoring and its Clinical Relevance for Rhythm Control Strategies

In some areas of clinical practice, it is not essential to have a continuous rhythm monitoring established. For example, in patients with oral anticoagulation and a rhythm control strategy, it is not of major importance whether they may have a few silent recurrences or not. Therapy in those patients is guided by symptomatic recurrences on the one hand, and information about basic AF characterisation in terms of a paroxysmal or persistent AF pattern on the other hand. Knowing the exact AF burden in addition to symptomatic recurrences is an interesting add-on but not crucial. In contrast, in some other patient cohorts, continuous and ‘gapless’ information about the rhythm status is mandatory. Patients who have an uncertain correlation of symptoms (possibly triggering significant therapy) to the underlying rhythm are objects for continuous monitoring. In addition, checking for silent AF in high-risk populations, possibly leading to the initiation of oral anticoagulation, is a wide area for future applications of continuous monitoring devices (see next section). In addition, in order to measure efficacy of approaches (e.g., rhythm control) and therapeutic techniques, it is more and more accepted that we need continuous rhythm information, because spotlight ECGs are missing a major part of the AF activity which may lead to a wrong evaluation of therapy efficacy.

Continuous Versus Intermittent Rhythm Monitoring in Transvenous Catheter Ablation of Atrial Fibrillation

The relevance of increasing the Holter observation length from 24 hours to seven days has been demonstrated by Kottkamp et al.33 One hundred patients underwent 24-hour and seven-day Holter monitoring post-pulmonary vein ablation for paroxysmal AF. At 12 months, ablation success rate was 88 % when using the 24-hour Holter data, but only 74 % as indicated by the seven-day Holter. One might argue that this gap in capturing recurrences outside a Holter registration interval could be closed if an external event recorder would be used. However, Klemm and co-workers demonstrated that, in 80 post-ablation patients using transtelephonic ECG recordings (minimum one ECG per day and in case of symptoms suggesting AF), during 54 % of transmitted ECGs demonstrating AF patients were asymptomatic.8 In 11 % of all tracings, the patients indicated symptoms but demonstrated stable sinus rhythm on the tracing. In line with this finding, it was demonstrated that, outside a blanking period of three months, the ablation success in patients with only symptomatic but ECG-documented recurrences was around 70 %, whereas, counting all ECG-documented AF recurrences using a tele-ECG concept, the percentage of patients with no AF recurrences went down to almost 40 %.35 It was then Pokushalov and co-workers who reported their wide and routine use of ICMs to perform long-term continuous AF detection in their AF ablation patients. Furthermore, they differentiated between sudden-onset and triggered-onset AF recurrence to individualise the therapeutical regimens for AF recurrences after the first catheter ablation.35,36 An AF burden of >4.5 % measured with continuous rhythm monitoring during the two months post-ablation blanking period was a powerful predictor of ablation failure. This information and correlation could trigger early re-intervention in those patients, shortening the period of time until stable sinus rhythm is reached.2

Continuous Versus Intermittent Rhythm Monitoring in Surgical Ablation of Atrial Fibrillation

Recent data indicate that, similar to catheter ablation, with surgical ablation, there is a significant proportion of patients with silent and intermittent AF recurrences, whereas a significant number of symptomatic AF episodes are not related to AF. Ip and co-workers report that in about 45 AF patients, who underwent video-assisted epicardial ablation and ICM implantation, as much as 46 % of the AF recurrences were asymptomatic, whereas only 66 % of the symptomatic episodes were AF-related.37

Bogachev-Prokophiev et al. reported ablation results for AF concomitant with mitral valve surgery after one year of continuous monitoring.38 Forty-seven patients with mitral valve disease and long-standing persistent AF underwent a left atrial maze procedure with bipolar radiofrequency ablation and valve surgery. The follow-up data recorded by an ICM were analysed after three, six and 12 months. On discharge, 40 (85.1 %) patients were in stable sinus rhythm as documented by in-clinic ECG; 4 (8.5 %) were in pacemaker rhythm; and 3 (6.4 %) were in AF. One (2.1 %) patient died after seven months. Upon 12-month follow-up examination, 30 (65.2 %) patients had an AF burden <0.5 % and were classified as responders. Three (6.5 %) of the 16 non-responders had atrial flutter and 13 (27.7 %) had documented AF recurrences with an AF burden >0.5 %. Two (4.3 %) patients with AF recurrences were completely asymptomatic. Among the symptomatic events stored by the patients, only 27.6 % were confirmed as genuine AF recurrences according to the concomitant ECG recorded by the ICM.38 Again, the results in these surgical AF patients demonstrate the importance of continuous rhythm monitoring, not only to verify the true amount of successful ablation therapy, but also to identify asymptomatic recurrences and to better discriminate between AF- and non-AF-related symptoms.

Atrial Fibrillation Detection and its Relevance to Oral Anticoagulation Therapy for Stroke Prophylaxis

AF has been associated with a fivefold increase in the risk of ischaemic stroke.39 When the risk of stroke is consistent, oral anticoagulation (OAC), with vitamin K antagonists or novel anticoagulants, is required for effective prevention of thromboembolic events. However, underutilisation of OAC therapy with warfarin has been shown to occur in ‘real world’ clinical practice in patients at moderate to high thromboembolic risk.40,41 Current guidelines for stroke prevention in AF39 do not differentiate between patients with symptomatic or asymptomatic AF, or between paroxysmal or persistent AF. The vast majority of AF episodes are asymptomatic, including many episodes of clinically significant duration.42 Implanted devices can provide continuous information on atrial rhythm as the basis for more accurate and extended diagnostic capabilities. On the basis of data provided by device diagnostics, the term ‘AF burden’ has been proposed as the total amount of time spent in AF per monitored time period and a duration of >5 minutes has been associated with stroke and death.43–46

Hazard Ratios from Studies

Article image

In summary, a series of studies performed in pacemaker or ICD patients indicate that longer times spent in AF are associated with slightly higher stroke risk, and that this increased risk is additive to well known stroke risk factors.43–46 A recent study performed on pacemaker patients assessed non only the CHADS2 score but also the CHA2DS2-VASc score, and found that implementation of device data on AF presence/duration/burden has the potential to contribute to improved clinical risk stratification.47

Recently, the ASSERT (Asymptomatic atrial fibrillation and stroke evaluation in pacemaker patients and the atrial fibrillation reduction atrial pacing trial) study included 2,580 patients aged 65 years or older, with hypertension and no history of AF, in whom a pacemaker or ICD had recently been implanted.14 Subclinical ATs (episodes of atrial rate >190 beats per minute for more than 6 minutes) occurred in 10 % of patients in the first three months and were associated with an increased risk of clinical AF (hazard ratio [HR] 5.56) and of ischaemic stroke or systemic embolism (HR 2.49). The conclusions of ASSERT were that subclinical ATs occur frequently in patients with an implanted device and are associated with a significantly increased risk of ischaemic stroke or systemic embolism. Table 1 shows the HRs from studies using different durations of AT or AF episodes.

The open question is: how often is the detailed and precise information on AF burden provided by devices (pacemakers, ICDs or ICMs) actually translated into clinical decision-making for the optimisation of thromboembolic prophylaxis? The ANGELS of AF (Anticoagulation use evaluation and life threatening events sentinels) project was a medical care programme aiming to improve OAC use as thromboprophylaxis, with the use of the information from the ICD AF diagnostics.48 The results demonstrate that a medical care programme may improve OAC implementation in standard medical practice by supplying accurate and continuous information about stroke risk factors, AF occurrence and burden.

In detail, 50 Italian cardiology clinics followed 3,438 ICD patients. In a subgroup of 15 centres (the ANGELS of AF centres), cardiologists attending follow-up visits were supplied with specific reports describing stroke risk factors and risk scores, including CHADS2, AF occurrence and duration, and current antithrombotic therapy for patients with AF (especially those with CHADS2 score >0 and without OAC therapy). The remaining centres represented a control group of patients as a comparison for OAC use. In the 15 ANGELS of AF centres, 36 % patients had AF and 29 % were not taking OAC. Appropriate OAC therapy was prescribed in 10 % of patients after evaluation of Angels of AF reports. The percentage of patients on OAC therapy, as indicated by guidelines, increased during follow-up from 46.1 % at baseline to 69.4 % at stroke risk evaluation phase, and up to 72.6 % at the end of the observation period. In control centres, corresponding figures were 46.9 % at baseline and 56.8 % at the end of the observation period (p<0.001 versus Angels AF group).

Since disparities in the use of guideline-recommended and evidence-based therapies for AF in inpatient and outpatient care settings are well documented, it is well known that a major underutilisation of OAC therapy in particular has been observed in numerous studies evaluating clinical practice.40 As a result, many AF patients may suffer embolic events, such as stroke or transient ischaemic attacks, which might have been prevented by appropriate detection through continuous monitoring of AF. The ANGELS of AF project demonstrates the possibility to improve OAC use in accordance with available guidelines for stroke risk reduction in AF by supplying attending physicians with reports about patients risk factors and AF information from continuous device monitoring.

A more focused device monitoring will be tested in CRYSTAL AF (Study of continuous cardiac monitoring to assess atrial fibrillation after cryptogenic stroke), a randomised prospective trial to evaluate a novel approach aimed at comparing standard arrhythmia monitoring (control arm) with implantation of a subcutaneous ICM (Reveal® XT, Medtronic, Inc., Minneapolis, Minnesota, US) (continuous monitoring arm) to detect AF in cryptogenic stroke patients.

Clinical Consequences of Atrial Fibrillation Monitoring in Implantable Cardioverter Defibrillator Therapy and Cardiac Resynchronisation Therapy

Avoidance of high-energy shocks remains an important goal in the management of patients with ICDs.16 Patients with inappropriate shocks more commonly have a history of AF, smoking and/or diastolic hypertension, and are more likely to also have had a prior appropriate ICD shock. Other investigators have found that a prior history of AF and appropriate therapy were predictors of inappropriate shocks.49 Because AF was the most common inappropriate shock mechanism, its association as a predictive factor is expected. Smoking was recently found to increase the incidence of both appropriate and inappropriate shocks in the MADIT (Multicenter automatic defibrillator implantation trial) II study, possibly because of a myriad of adverse consequences such as sympathetic stimulation, increased platelet reactivity, vasoconstriction, endothelial dysfunction and tachycardia.50 Hypertension is a potent risk factor for AF and may act in this way to promote the likelihood of an inappropriate shock. The link between appropriate and inappropriate therapy likely stems from a combination of several factors:

  • ventricular arrhythmia or its treatment (antitachycardia pacing techniques or shock) provoking AF;51,52
  • inappropriate therapy causing ventricular tachycardia (VT); that is, pro-arrhythmia;53
  • a common factor or factors predisposing to both VT/ventricular fibrillation and AF or supraventricular tachycardia; or
  • incorrect categorisation of some appropriate episodes in a given patient as inappropriate or vice versa.54

Numerous trials have identified a rate of inappropriate shocks between 20 % and 30 %, most commonly due to rapidly conducted AF. It was demonstrated that poor rate control often occurs days and even months before shocks, and therefore continuous monitoring along with the use of patient or wireless alerts may be warranted. The use of continuous monitoring along with automated alerts to provide early notification of AF with a fast ventricular rate may help clinicians take action early in order to prevent future shocks.16

There has been increasing recognition of the epidemics of heart failure (HF) and AF, conditions often present in the same patient. Initially validated in populations in normal sinus rhythm, CRT now has an established role in the treatment of HF patients with left ventricular (LV) dysfunction and intraventricular conduction delay. The incidence of new AF in patients who have been implanted with a CRT device is substantial, exceeding one-fourth of patients.

It is clear that CRT is not beneficial if biventricular (BiV) pacing does not reliably occur. This is because the physiological impact of CRT is postulated to occur through synchronising ventricular contraction, leading to improvement in LV filling and pumping efficiency and reduction in the extent of functional mitral regurgitation. In patients with HF and AF, AV synchrony does not exist; thus, any clinical benefit is predicated on BiV synchronisation. However, in patients with AF, erratic electric activity of the atria occasionally will penetrate the AV node and override, interrupt or disrupt BiV capture.55,56 This leads to fusion or pseudo-fusion beats and suboptimal CRT delivery, particularly in situations of increased myocardial demand, as occurs from increased adrenergic tone during stress or exercise. Perhaps even more problematic than the lack of BiV capture is that the percentage of BiV pacing recorded by internal CRT device counters is often artificially inflated during AF, due to erroneous counting of fusion and pseudo-fusion complexes.57 Although a clinician might believe that adequate delivery of CRT has occurred because of an apparently high percentage of BiV pacing, a patient with AF may not have received the full benefit of synchronised BiV capture.58

To rely on device counters alone can be misleading and to review data from the device diagnostics or to optimise the AF detection is instrumental to ensure that pseudo-fused and fused beats are not being miscounted as those associated with BiV capture.

Santini et al. have monitored AF in 1,193 HF patients with a CRT-D (CRT defibrillator) and found that 30 % had AF episodes lasting longer than 10 minutes.59 Compared with the group that had only sinus rhythm, the group with AF episodes had a 2.16 times higher risk of mortality or HF hospitalisation. The authors concluded that AF detection in CRT-D patients is important to identify patients at risk of cardiac deterioration or patients with suboptimal rate or rhythm control.

Conclusion

We have reviewed the major areas of clinical indication for AF monitoring. For outcome measurements in AF treatment trials comparing rhythm versus rate control or different rhythms strategies continuos AF monitoring is gaining more and more importance. It is quite obvious that intermittent ECG documentation only captures a minority of the true AF burden. Whereas, in the pure clinical arena especially in the field of rhythm control correlation of symptoms and underlying rhythm is still more relevant than capturing different levels of AF burden.

The method for AF monitoring which should be used has to be tailored to the patients situation; implantable long term ECG monitors offer a lot of advantages but require the willingness of an invasive approach.

Detecting AF in patients with risk factors should be performed wherever it is possible, especially when we have implanted cardiac rhythm devices. We know that even short lasting AF episodes with a low AF burden already induce a significant risk increase, therefore an electrogram-documented and -verified device-based AF detection should trigger the process of starting OAC as recommended for every patient with risk factors and surface ECG documented AF. In addition, device-based AF detection has importance for device therapy itself like device programming, e.g. to prevent inadequate shock therapy from rapidly conducted AF in defibrillator patients.

One step further; searching for AF in high risk populations with so far not-documented AF – like the cryptogenic stroke population – is subject of scientific studies and has the potential for enlarging the indication for AF monitoring significantly.

Termination of OAC in patients with risk factors and AF is not recommended; however in selected cases with manifest bleeding or an excessive risk for bleeding it may be necessary to execute a bail-out concept with either left atrial occluder implantation or rhythm control with continuous monitoring and triggered OAC in case of arrhythmia recurrence.

References

  1. Mittal S, Movsowitz C, Steinberg JS. Ambulatory external electrocardiographic monitoring: focus on atrial fibrillation. J Am Coll Cardiol 2011;58(17):1741–9.
    Crossref | PubMed
  2. Pokushalov E, Romanov A, Corbucci G, et al. Does atrial fibrillation burden measured by continuous monitoring during the blanking period predict the response to ablation at 12- month follow-up?. Heart Rhythm 2012; [Epub ahead of print].
  3. Patten M, Maas R, Karim A, et al. Event-recorder monitoring in the diagnosis of atrial fibrillation in symptomatic patients: subanalysis of the SOPAT trial. J Cardiovasc Electrophysiol 2006;17(11):1216–20.
    Crossref | PubMed
  4. Fetsch T, Bauer P, Engberding R, et al. Prevention of atrial fibrillation after cardioversion: results of the PAFAC trial. Eur Heart J 2004;25(16):1385–94.
    Crossref | PubMed
  5. Nergårdh A, Frick M. Perceived heart rhythm in relation to ECG findings after direct current cardioversion of atrial fibrillation. Heart 2006;92(9):1244–7.
    Crossref | PubMed
  6. Hindricks G, Piorkowski C, Tanner H, et al. Perception of atrial fibrillation before and after radiofrequency catheter ablation: relevance of asymptomatic arrhythmia recurrence. Circulation 2005;112(3):307–13.
    Crossref | PubMed
  7. Senatore G, Stabile G, Bertaglia E, et al. Role of transtelephonic electrocardiographic monitoring in detecting short-term arrhythmia recurrences after radiofrequency ablation in patients with atrial fibrillation. J Am Coll Cardiol 2005;45(6):873–6.
    Crossref | PubMed
  8. Klemm HU, Ventura R, Rostock T, et al. Correlation of symptoms to ECG diagnosis following atrial fibrillation ablation. J Cardiovasc Electrophysiol 2006;17(2):146–50.
    Crossref | PubMed
  9. Vasamreddy CR, Dalal D, Dong J, et al. Symptomatic and asymptomatic atrial fibrillation in patients undergoing radiofrequency catheter ablation. J Cardiovasc Electrophysiol 2006;17(2):134–9.
    Crossref | PubMed
  10. Strickberger SA, Ip J, Saksena S, et al. Relationship between atrial tachyarrhythmias and symptoms. Heart Rhythm 2005;2(2):125–31.
    Crossref | PubMed
  11. Israel CW, Grönefeld G, Ehrlich JR, et al. Long-term risk of recurrent atrial fibrillation as documented by an implantable monitoring device: implications for optimal patient care. J Am Coll Cardiol 2004;43(1):47–52.
    Crossref | PubMed
  12. Israel CW, Neubauer H, Olbrich HG, et al. Incidence of atrial tachyarrhythmias in pacemaker patients: results from the Balanced Evaluation of Atrial Tachyarrhythmias in Stimulated patients (BEATS) study. Pacing Clin Electrophysiol 2006;29(6):582–8.
    Crossref | PubMed
  13. Orlov MV, Ghali JK, Araghi-Niknam M, et al. Asymptomatic atrial fibrillation in pacemaker recipients: incidence, progression, and determinants based on the atrial high rate trial. Pacing Clin Electrophysiol 2007;30(3):404–11.
    Crossref | PubMed
  14. Healey JS, Connolly SJ, Gold MR, et al. Subclinical atrial fibrillation and the risk of stroke. N Engl J Med 2012;366(2):120–9.
    Crossref | PubMed
  15. Sinha AM, Diener HC, Morillo CA, et al. Cryptogenic Stroke and underlying Atrial Fibrillation (CRYSTAL AF): design and rationale. Am Heart J 2010;160(1):36–41e1.
    Crossref | PubMed
  16. Fischer A, Ousdigian KT, Johnson JW, et al. The impact of atrial fibrillation with rapid ventricular rates and device programming on shocks in 106,513 ICD and CRT-D patients. Heart Rhythm 2012;9(1):24–31.
    Crossref | PubMed
  17. Francis DA, Heron JR, Clarke M. Ambulatory electrocardiographic monitoring in patients with transient focal cerebral ischaemia. J Neurol Neurosurg Psychiatry 1984;47(3):256–9.
    Crossref | PubMed
  18. Koudstaal PJ, van Gijn J, Klootwijk AP, et al. Holter monitoring in patients with transient and focal ischemic attacks of the brain. Stroke 1986;17(2):192–5.
    Crossref | PubMed
  19. Kessler DK, Kessler KM. Is ambulatory electrocardiography useful in the evaluation of patients with recent stroke?. Chest 1995;107(4):916–8.
    Crossref | PubMed
  20. Abdon NJ, Zettervall O, Carlson J, et al. Is occult atrial disorder a frequent cause of non-hemorrhagic stroke? Long-term ECG in 86 patients. Stroke 1982;13(6):832–7.
    Crossref | PubMed
  21. Schuchert A, Behrens G, Meinertz T. Impact of long-term ECG recording on the detection of paroxysmal atrial fibrillation in patients after an acute ischemic stroke. Pacing Clin Electrophysiol 1999;22(7):1082–4.
    Crossref | PubMed
  22. Roche F, Gaspoz JM, Da Costa A, et al. Frequent and prolonged asymptomatic episodes of paroxysmal atrial fibrillation revealed by automatic long-term event recorders in patients with a negative 24-hour Holter. Pacing Clin Electrophysiol 2002;25(11):1587–93.
    Crossref | PubMed
  23. Kinlay S, Leitch JW, Neil A, et al. Cardiac event recorders yield more diagnoses and are more cost-effective than 48-hour Holter monitoring in patients with palpitations. A controlled clinical trial. Ann Intern Med 1996;124(1 Pt 1):16–20.
    Crossref | PubMed
  24. Roche F, Gaspoz JM, Pichot V, et al. Accuracy of an automatic and patient-triggered long-term solid memory ambulatory cardiac event recorder. Am J Cardiol 1997;80(8):1095–8.
    Crossref | PubMed
  25. Camm AJ, Lip GY, De Caterina R, et al. 2012 focused update of the ESC Guidelines for the management of atrial fibrillation. An update of the 2010 ESC Guidelines for the management of atrial fibrillation. Eur Heart J 2012; [Epub ahead of print].
  26. Pollak WM, Simmons JD, Interian A Jr, et al. Clinical utility of intraatrial pacemaker stored electrograms to diagnose atrial fibrillation and flutter. Pacing Clin Electrophysiol 2001;24(4 Pt 1):424–9.
    Crossref | PubMed
  27. Hoffmann E, Sulke N, Edvardsson N, et al. New insights into the initiation of atrial fibrillation: a detailed intraindividual and interindividual analysis of the spontaneous onset of atrial fibrillation using new diagnostic pacemaker features. Circulation 2006;113(16):1933–41.
    Crossref | PubMed
  28. Perez FJ, Lung TH, Ellenbogen KA, Wood MA. Is time to first recurrence of atrial fibrillation correlated with atrial fibrillation burden?. Am J Cardiol 2006;97(9):1343–5.
    Crossref | PubMed
  29. Padeletti L, Botto G, Spampinato A, et al. Prevention of paroxysmal atrial fibrillation in patients with sinus bradycardia: role of right atrial linear ablation and pacing site. J Cardiovasc Electrophysiol 2003;14(7):733–8.
    Crossref | PubMed
  30. Lee MA, Weachter R, Pollak S, et al. The effect of atrial pacing therapies on atrial tachyarrhythmia burden and frequency: results of a randomized trial in patients with bradycardia and atrial tachyarrhythmias. J Am Coll Cardiol 2003;41(11):1926–32.
    Crossref | PubMed
  31. Hindricks G, Pokushalov E, Urban L, et al. Performance of a new leadless implantable cardiac monitor in detecting and quantifying atrial fibrillation: Results of the XPECT trial. Circ Arrhythm Electrophysiol 2010;3(2):141–7.
    Crossref | PubMed
  32. Camm AJ, Corbucci G, Padeletti L. Usefulness of continuous electrocardiographic monitoring for atrial fibrillation. Am J Cardiol 2012;110(2):270–6.
    Crossref | PubMed
  33. Kottkamp H, Tanner H, Kobza R, et al. Time courses and quantitative analysis of atrial fibrillation episode number and duration after circular plus linear left atrial lesions: trigger elimination or substrate modification: early or delayed cure?. J Am Coll Cardiol 2004;44(4):869–77.
    Crossref | PubMed
  34. Piorkowski C, Kottkamp H, Tanner H, et al. Value of different follow-up strategies to assess the efficacy of circumferential pulmonary vein ablation for the curative treatment of atrial fibrillation. J Cardiovasc Electrophysiol 2005;16(12):1286–92.
    Crossref | PubMed
  35. Pokushalov E, Romanov A, Corbucci G, et al. Ablation of paroxysmal and persistent atrial fibrillation: 1-year follow-up through continuous subcutaneous monitoring. J Cardiovasc Electrophysiol. 2011;22(4):369–75.
    Crossref | PubMed
  36. Pokushalov E, Romanov A, Corbucci G, et al. Use of an implantable monitor to detect arrhythmia recurrences and select patients for early repeat catheter ablation for atrial fibrillation: a pilot study. Circ Arrhythm Electrophysiol 2011;4(6):823–31.
    Crossref | PubMed
  37. Ip JH, Viqar-Syed M, Grimes D, et al. Surveillance of AF recurrence post-surgical AF ablation using implantable cardiac monitor. J Interv Card Electrophysiol 2012;33(1):77–83.
    Crossref | PubMed
  38. Bogachev-Prokophiev A, Zheleznev S, Romanov A, et al. Ablation for atrial fibrillation during mitral valve surgery: 1-year results through continuous subcutaneous monitoring. Interact Cardiovasc Thorac Surg 2012;15(1):37–41.
    Crossref | PubMed
  39. European Heart Rhythm A, European Association for Cardio- Thoracic S, Camm AJ, Kirchhof P, Lip GY, et al. Guidelines for the management of atrial fibrillation: the Task Force for the Management of Atrial Fibrillation of the European Society of Cardiology (ESC). Europace 2010;12(10):1360–420.
    Crossref | PubMed
  40. Nieuwlaat R, Capucci A, Lip GY, et al. Antithrombotic treatment in real-life atrial fibrillation patients: a report from the Euro Heart Survey on Atrial Fibrillation. Eur Heart J 2006;27(24):3018–26.
    Crossref | PubMed
  41. Nieuwlaat R, Olsson SB, Lip GY, et al. Guideline-adherent antithrombotic treatment is associated with improved outcomes compared with undertreatment in high-risk patients with atrial fibrillation. The Euro Heart Survey on Atrial Fibrillation. Am Heart J 2007;153(6):1006–12.
    Crossref | PubMed
  42. Savelieva I, Camm AJ. Clinical relevance of silent atrial fibrillation: prevalence, prognosis, quality of life, and management. J Interv Card Electrophysiol 2000;4(2):369–82.
    PubMed
  43. Glotzer TV, Hellkamp AS, Zimmerman J, et al. Atrial high rate episodes detected by pacemaker diagnostics predict death and stroke: report of the Atrial Diagnostics Ancillary Study of the MOde Selection Trial (MOST). Circulation 2003;107(12):1614–9.
    Crossref | PubMed
  44. Capucci A, Santini M, Padeletti L, et al. Monitored atrial fibrillation duration predicts arterial embolic events in patients suffering from bradycardia and atrial fibrillation implanted with antitachycardia pacemakers. J Am Coll Cardiol 2005;46(10):1913–20.
    Crossref | PubMed
  45. Botto GL, Padeletti L, Santini M, et al. Presence and duration of atrial fibrillation detected by continuous monitoring: crucial implications for the risk of thromboembolic events. J Cardiovasc Electrophysiol 2009;20(3):241–8.
    Crossref | PubMed
  46. Glotzer TV, Daoud EG, Wyse DG, et al. The relationship between daily atrial tachyarrhythmia burden from implantable device diagnostics and stroke risk: the TRENDS study. Circ Arrhythm Electrophysiol 2009;2(5):474–80.
    Crossref | PubMed
  47. Boriani G, Botto GL, Padeletti L, et al. Improving stroke risk stratification using the CHADS2 and CHA2DS2-VASc risk scores in patients with paroxysmal atrial fibrillation by continuous arrhythmia burden monitoring. Stroke 2011;42(6):1768–70.
    Crossref | PubMed
  48. Boriani G, Santini M, Lunati M, et al. Improving thromboprophylaxis using atrial fibrillation diagnostic capabilities in implantable cardioverter-defibrillators: the multicentre Italian ANGELS of AF Project. Circ Cardiovasc Qual Outcomes 2012;5(2):182–8.
    Crossref | PubMed
  49. Theuns DA, Klootwijk AP, Simoons ML, Jordaens LJ. Clinical variables predicting inappropriate use of implantable cardioverter-defibrillator in patients with coronary heart disease or nonischemic dilated cardiomyopathy. Am J Cardiol 2005;95(2):271–4.
    Crossref | PubMed
  50. Goldenberg I, Moss AJ, McNitt S, et al. Cigarette smoking and the risk of supraventricular and ventricular tachyarrhythmias in high-risk cardiac patients with implantable cardioverter defibrillators. J Cardiovasc Electrophysiol 2006;17(9):931–6.
    Crossref | PubMed
  51. Johnson NJ, Marchlinski FE. Arrhythmias induced by device antitachycardia therapy due to diagnostic nonspecificity. J Am Coll Cardiol 1991;18(5):1418–25.
    Crossref | PubMed
  52. Florin TJ, Weiss DN, Peters RW, et al. Induction of atrial fibrillation with low-energy defibrillator shocks in patients with implantable cardioverter defibrillators. Am J Cardiol 1997;80(7):960–2.
    Crossref | PubMed
  53. Alter P, Waldhans S, Plachta E, et al. Complications of implantable cardioverter defibrillator therapy in 440 consecutive patients. Pacing Clin Electrophysiol 2005;28(9):926–32.
    Crossref | PubMed
  54. Daubert JP, Zareba W, Cannom DS, et al. Inappropriate implantable cardioverter-defibrillator shocks in MADIT II: frequency, mechanisms, predictors, and survival impact. Am Coll Cardiol 2008;51(14):1357–65.
    Crossref | PubMed
  55. Upadhyay GA, Choudhry NK, Auricchio A, et al. Cardiac resynchronization in patients with atrial fibrillation: a meta-analysis of prospective cohort studies. J Am Coll Cardiol 2008;52(15):1239–46.
    Crossref | PubMed
  56. Gasparini M, Auricchio A, Regoli F, et al. Four-year efficacy of cardiac resynchronization therapy on exercise tolerance and disease progression: the importance of performing atrioventricular junction ablation in patients with atrial fibrillation. J Am Coll Cardiol 2006;48(4):734–43.
    Crossref | PubMed
  57. Kamath GS, Cotiga D, Koneru JN, et al. The utility of 12-lead Holter monitoring in patients with permanent atrial fibrillation for the identification of nonresponders after cardiac resynchronization therapy. J Am Coll Cardiol 2009;53(12):1050–5.
    Crossref | PubMed
  58. Upadhyay GA, Steinberg JS. Managing atrial fibrillation in the CRT patient: Controversy or consensus?. Heart Rhythm 2012;9(8 Suppl.):S51–9.
    Crossref | PubMed
  59. Santini M, Gasparini M, Landolina M, et al. Device-detected atrial tachyarrhythmias predict adverse outcome in real-world patients with implantable biventricular defibrillators. J Am Coll Cardiol 2011;57:167–72.
    Crossref | PubMed